A bifurcated fiber optic bundle brings two independently routed fiber groups to a single polished face. The common-end drawing determines where each branch delivers or collects light.
Specify the common ferrule, the fiber groups visible at its face, and the route to each labeled branch. Include fiber type, core size, numerical aperture, wavelength band, branch lengths, terminations and a per-path optical test. The fiber count in each branch describes the routing. Measured throughput depends on launch, collection and end-face conditions.
How a bifurcated fiber optic bundle routes light
At the common end, individual fiber ends sit in one ferrule or probe head. The fibers separate into branch A and branch B farther along the assembly. A reflection probe can use one branch to illuminate a sample and the other to carry the returned signal to a spectrometer. The Avantes reflection probe documentation illustrates this configuration with an illumination group and a central read fiber.

| Assembly | Design decision | What to put on the drawing |
|---|---|---|
| Bifurcated bundle | Two fiber groups share a polished face | Common-end geometry, fiber groups and the branch map |
| Connectorized fan-out | Individual fibers reach labeled ports | Channel-to-port map, breakout lengths and connector details |
| Optical splitter or coupler | Input optical power is distributed among output ports | Split ratio, wavelength band, insertion loss and uniformity as applicable |
The fiber optic bundle selection guide covers bundle construction and termination options. If the drawing calls for separately terminated ports, review the breakout, fan-out and pigtail comparison.
Draw the common face before choosing the branches
Define the viewing direction into the polished face, then show an orientation datum such as a key or flat where needed. An interleaved layout distributes both groups over the active area. A separated layout puts each branch in a designated region. A central collection group with illumination around it is another option when the optics require it. Choose the arrangement using the instrument aperture, spot and field of view.

- Fiber and light. Specify material, core and cladding dimensions, numerical aperture, wavelength band and launch conditions.
- Ferrule. Define outside diameter, active area, setback or protrusion, polish and alignment datum. Where geometry affects coupling, state allowed inactive spacing or fill requirements.
- Branches. Dimension the overall and branch lengths from named datums. Identify A/B permanently and define sheath, strain relief, end fittings and routing limits.
Example for a two-branch reflection head
Consider a conceptual common end with 12 illumination fibers routed to branch A and 24 collection fibers routed to branch B. The equipment drawing would define where the 12 and 24 fibers sit in the polished face, the active area, working geometry, and each branch’s termination. This is an illustrative fiber count for drafting a specification.
Test A by launching into the source-side branch and measuring at the common face with agreed collection optics. Test B by applying the agreed reference illumination at the common face and measuring at the detector-side branch. Record wavelength, reference method and setup geometry for both tests. The 12:24 fiber count alone cannot predict either measured throughput.
Agree acceptance before requesting a build
- Confirm the face drawing and branch labels against the released revision.
- Verify each branch map with controlled illumination and inspect the polished ends using the agreed inspection method.
- Measure per-path throughput or loss with the agreed launch, collection geometry and reference planes. Document the setup so that the receiving test can reproduce it.
- Check branch length, bend envelope, termination fit and any project-specific environmental requirements.
Frequently asked questions
Is a bifurcated bundle the same as a 1×2 splitter?
A bifurcated bundle routes physical fiber groups to two branches. A 1×2 splitter divides optical power from an input into two output ports. Specify which behavior the instrument requires.
Can illumination and collection use different fiber types?
Ask for a feasibility review using the required core sizes, numerical apertures, wavelength bands and ferrule geometry. The approved assembly drawing should identify the fiber type in each group.
How should branch lengths be measured?
Mark the split point and end reference planes on the drawing. State the length and tolerance for each branch and the overall path, including how connector boots or strain relief enter the measurements.
Request a bifurcated bundle quote
Send the common-end sketch, A/B fiber map, wavelength band, ferrule and branch dimensions, termination details, quantity and per-path acceptance targets.